Journals / Cyprus Journal of Medical Sciences / 2020 / Cilt: 5 - Sayı: 4

Neonates Hearing Screening Results: A Comparison of Chirp and Click Stimuli with an Automated Auditory Brainstem Response Device

Pages
313–316
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

CONCLUSION We were unable to prove that the chirp stimulus could be replaced by the click stimuli in neonatal hearing screening using AABR device.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:BACKGROUND/AIMS Automated Auditory Brain Response (AABR) devices are unique tools for neonate hearing screening programs.
📊
Key Findings & Data:The most common risk factors were consanguineous marriage of parents (n=184), history of being admitted to the neonatal intensive care unit (n=119), and jaundice (n=102).
💡
Conclusion & Impact:CONCLUSION We were unable to prove that the chirp stimulus could be replaced by the click stimuli in neonatal hearing screening using AABR device.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

BACKGROUND/AIMS Automated Auditory Brain Response (AABR) devices are unique tools for neonate hearing screening programs. Contemporary utilization of AABR devices is based on two stimuli called chirp and click. The first study objective was to identify any possible differences in chirp and click stimulus results of the AABR devices in neonates. The second aim was to investigate any possible risk factor that could affect the results of each stimulus. MATERIAL and METHODS Chirp and click stimuli were applied to each neonate using the AABR devices. Results were recorded automatically as pass or failed. Those with failed results were called after month for a retest. Hearing loss risk factors were obtained from the parents and caregivers. RESULTS Twenty-one of the fifty-eight chirp stimuli gave negative results on the AABR in the second AABR test. Twenty-nine of the seventy-six click stimuli on the AABR gave a failed result on retest. The most common risk factors were consanguineous marriage of parents (n=184), history of being admitted to the neonatal intensive care unit (n=119), and jaundice (n=102). In addition, hearing loss was not detected in any neonate and was therefore not considered to significantly affect the results of the chirp or click stimulus on the AABR. CONCLUSION We were unable to prove that the chirp stimulus could be replaced by the click stimuli in neonatal hearing screening using AABR device.

🔖

Cite This Article

DENİZ, H., YAZICI, A., ÇOŞKUN, M. E. (2020). Neonates Hearing Screening Results: A Comparison of Chirp and Click Stimuli with an Automated Auditory Brainstem Response Device. Cyprus Journal of Medical Sciences, Cilt: 5 - Sayı: 4, 313–316. https://ajindex.com/articles/216010

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (11)

  1. Erenberg A, Lemons J, Sia C, Trunkel D, Ziring P, Adams M, et al. Newborn and infant hearing loss: detection and intervention. American Academy of Pediatrics. Task Force on Newborn and Infant Hearing, 1998- 1999. Pediatrics 1999
  2. 103(2): 527-30. [Crossref] 2. Newborn and Infant Hearing Screening CURRENT ISSUES AND GUIDING PRINCIPLES FOR ACTION WHO Library Cataloguing-in-Publication Data Newborn and Infant Hearing Screening: Current Issues and Guiding Principles for Action. https://www.who.int/blindness/publications/Newborn_and_Infant_Hearing_Screening_Report.pdf. Accessed August 7, 2019. 3. Hahn M, Lamprecht-Dinnesen A, Heinecke A, Hartmann S, Bülbül S, Schröder G, et al. Hearing screening in healthy newborns: feasibility of different methods with regard to test time. Int J Pediatr Otorhinolaryngol 1999
  3. 51(2): 83-9. [Crossref] 4. Iwasaki S, Hayashi Y, Seki A, Nagura M, Hashimato Y, Oshima G, et al. A model of two-stage newborn hearing screening with auto mated auditory brainstem response. Int J Pediatr Otorhinolaryngol 2003
  4. 67(10): 1099-104. [Crossref] 5. Dau T, Wegner O, Mellert V, Kollmeier B. Auditory brainstem responses with optimized chirp signals compensating basilar-membrane dispersion. J Acoust Soc Am 2000
  5. 107(3): 1530-40. [Crossref] 6. Zirn S, Louza J, Reiman V, Wittlinger N, Hempel JM, Schuster M. Comparison between ABR with click and narrow band chirp stimuli in children. Int J Pediatr Otorhinolaryngol 2014
  6. 78(8): 1352-5. [Crossref] 7. Venail F, Artaud JP, Blanchet C, Uziel A, Mondain M. Refining the audiological assessment in children using narrow-band CE-Chirp-evoked auditory steady state responses. Int J Audiol 2015
  7. 54(2): 106-13. [Crossref] 8. Cebulla M, Lurz H, Shehata-Dieler W. Evaluation of waveform, latency and amplitude values of chirp ABR in newborns. Int J Pediatr Otorhinolaryngol 2014
  8. 78(4): 631-6. [Crossref] 9. Elberling C, Don M, Cebulla M, Stürzebecher E. Auditory steady-state responses to chirp stimuli based on cochlear traveling wave delay. J Acoust Soc Am 2007
  9. 122(5): 2772. [Crossref] 10. Maloff ES, Hood LJ. A comparison of auditory brain stem responses elicited by click and chirp stimuli in adults with normal hearing and sensory hearing loss. Ear Hear 2014
  10. 35(2): 271-82. [Crossref] 11. Cebulla M, Hofmann S, Shehata-Dieler W. Sensitivity of ABR based newborn screening with the MB11 BERAphone®. Int J Pediatr Otorhinolaryngol 2014
  11. 78(5): 756-61. [Crossref] 12. Levit Y, Himmelfarb M, Dollberg S. Sensitivity of the Auto