Objective & Hypothesis:PURPOSE Clinical cases of stent-fractures show that corrosion behavior might play a role in these fractures.
📊
Key Findings & Data:METHODSA total of 28 peripheral stents of different materials (nitinol, cobalt-chromium-nickel, tantalum, V4A) and surface treatments (electropolish, mechanical polish, no polish) were tested in vitro.
💡
Conclusion & Impact:CONCLUSIONThe analysis of corrosion behavior may be useful to select the right stent fulfilling the individual needs of the patient within a large number of different stents.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

PURPOSE Clinical cases of stent-fractures show that corrosion behavior might play a role in these fractures. Implanted in vivo, especially in combination with other implanted foreign materials, these metallic products are exposed to special conditions, which can cause a process of corrosion. Here, we aimed to test the corrosion potential of stents made of different materials in an in vitro setting. METHODSA total of 28 peripheral stents of different materials (nitinol, cobalt-chromium-nickel, tantalum, V4A) and surface treatments (electropolish, mechanical polish, no polish) were tested in vitro. Corrosion was accelerated by applying a constant voltage of 3.5 V and amperage of 1.16 mA in 0.9% NaCl. RESULTSNitinol stents showed the lowest susceptibility to corrosion and the longest period without dam-age. The Memotherm II® (BARD Angiomed®) was the only stent that showed neither macroscop-ic nor microscopic damages. The worst performing material was cobalt-chromium-nickel, which showed corrosion damages about ten times earlier compared to nitinol. Considering the reasons for termination of the test, nitinol stents primarily showed length deficits, while V4A and tantalum stents showed fractures. Cobalt-chromium-nickel stents had multiple fractures or a complete lysis in equal proportions. When placed in direct contact, nitinol stents showed best corrosion resis-tance, regardless of what material they were combined with. In terms of polishing treatments, elec-tropolished stents performed the best, mechanical-polished stents and those without polishing treatment followed. CONCLUSIONThe analysis of corrosion behavior may be useful to select the right stent fulfilling the individual needs of the patient within a large number of different stents.

🔖

Cite This Article

PAPROTTKA, J. K., PAPROTTKA, M. P., REİSER, F. M., WAGGERSHAUSER, T. (2015). Comparative study of the corrosion behavior of peripheral stents in an accelerated corrosion model: experimental in vitro study of 28 metallic vascular endoprostheses. Diagnostic and Interventional Radiology, Cilt: 21 - Sayı: 5, 403–409. https://ajindex.com/articles/213559

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (2)

  1. Meyer JA. Werner Forssmann and catheteriza- tion of the heart, 1929. Ann Thorac Surg 1990; 49:497-499. [CrossRef] Schatz RA. A view of vascular stents. Circulation 1989; 79:445-457. [CrossRef] Sniderman KW. Noncoronary vascular stenting. Prog Cardiovasc Dis 1996; 39:141-164. [CrossRef] Palmaz JC, Richter GM, Noeldge G, et al. Intra- luminal stents in atherosclerotic iliac artery ste- nosis: preliminary report of a multicenter study. Radiology 1988; 168:727-731. [CrossRef] Guthrie J, Battat B, Grethlein C. Accelerated cor- rosion testing. The AMPTIAC Quarterly 2002; 6. Shih CC, Lin SJ, Chen YL, et al. The cytotoxicity of corrosion products of nitinol stent wire on cultured smooth muscle cells. J Biomed Mater Res 2000; 52:395-403. [CrossRef] Shih CC, Shih CM, Chen YL, et al. Growth inhi- bition of cultured smooth muscle cells by cor- rosion products of 316 L stainless steel wire. J Biomed Mater Res 2001; 57:200-207. [CrossRef] Stepan LL, Levi DS, Gans E, Mohanchandra KP, Ujihara M, Carman GP. Biocorrosion investiga- tion of two shape memory nickel based alloys: Ni-Mn-Ga and thin film NiTi. J Biomed Mater Res A 2007; 82:768-776. [CrossRef] Wang J, Li J, Tang J, Lu S, Xi T. A long term accel- erating corrosion fatigue texting of coronary stents in vitro. Sheng Wu Yi Xue Gong Cheng Xue Za Zhi 2008; 25:398-401.
  2. Wiskirchen J, Venugopalan R, Holton AD, et al. Radiopaque markers in endovascular stents--benefit and potential hazards. Rofo 2003; 175:484-488. [CrossRef] Pound BG. Susceptibility of nitinol to local- ized corrosion. J Biomed Mater Res A 2006; 77:185-191. [CrossRef] Thierry B, Tabrizian M, Trepanier C, Savadogo O, Yahia L. Effect of surface treatment and ster- ilization processes on the corrosion behavior of NiTi shape memory alloy. J Biomed Mater Res 2000; 51:685-693. [CrossRef] Trepanier C, Tabrizian M, Yahia LH, Bilodeau L, Piron DL. Effect of modification of oxide layer on NiTi stent corrosion resistance. J Biomed Mater Res 1998; 43:433-440. [CrossRef] Venugopalan R. Corrosion testing of stents: a novel fixture to hold entire device in deployed form and finish. J Biomed Mater Res 1999; 48:829-832. [CrossRef] Endo K, Sachdeva R, Araki Y, Ohno H. Effects of titanium nitride coatings on surface and corro- sion characteristics of Ni-Ti alloy. Dent Mater J 1994; 13:228-239. [CrossRef] Speck KM, Fraker AC. Anodic polarization be- havior of Ti-Ni and Ti-6A1-4V in simulated phys- iological solutions. J Dent Res 1980; 59:1590- 1595. [CrossRef] Shabalovskaya S