Journals / Diagnostic and Interventional Radiology / 2015 / Cilt: 21 - Sayı: 6

Impact of radiation dose and iterative reconstruction on pulmonary nodule measurements at chest CT: a phantom study

Pages
459–465
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

However, for GGNs, IMR may help reduce diameter measurement error while improving image quality.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:PURPOSE We aimed to identify the impact of radiation dose and iterative reconstruction (IR) on measure-ment of pulmonary nodules by chest computed tomography (CT).METHODSCT scans were performed on a chest phantom containing various nodules (diameters of 3, 5, 8, 10, and 12 mm; +100, -630 and -800 HU for each diameter) at 80, 100, 120 kVp and 10, 20, 50, 100 mAs (a total of 12 radiation dose settings).
📊
Key Findings & Data:Each CT was reconstructed using filtered back projec-tion, iDose4, and iterative model reconstruction (IMR).
💡
Conclusion & Impact:However, for GGNs, IMR may help reduce diameter measurement error while improving image quality.
🔬Design:Modeling & Simulation
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

PURPOSE We aimed to identify the impact of radiation dose and iterative reconstruction (IR) on measure-ment of pulmonary nodules by chest computed tomography (CT).METHODSCT scans were performed on a chest phantom containing various nodules (diameters of 3, 5, 8, 10, and 12 mm; +100, -630 and -800 HU for each diameter) at 80, 100, 120 kVp and 10, 20, 50, 100 mAs (a total of 12 radiation dose settings). Each CT was reconstructed using filtered back projec-tion, iDose4, and iterative model reconstruction (IMR). Thereafter, two radiologists measured the diameter and attenuation of the nodules. Noise, contrast-to-noise ratio and signal-to-noise ratio of CT images were also obtained. Influence of radiation dose and reconstruction algorithm on mea-surement error and objective image quality metrics was analyzed using generalized estimating equations.RESULTSThe 80 kVp, 10 mAs CT scan was not feasible for the measurement of 3 mm sized simulated ground-glass nodule (GGN); otherwise, diameter measurement error was not significantly influenced by radiation dose (P > 0.05). IR did not have a significant impact on diameter measurement error for simulated solid nodules (P > 0.05). However, for simulated GGNs, IMR was associated with signifi-cantly decreased relative diameter measurement error (P < 0.001). Attenuation measurement error was not significantly influenced by either radiation dose or reconstruction algorithm (P > 0.05). Objective image quality was significantly better with IMR (P < 0.05).CONCLUSIONNodule measurements were not affected by radiation dose except for 3 mm simulated GGN on 80 kVp, 10 mAs dose setting. However, for GGNs, IMR may help reduce diameter measurement error while improving image quality.

🔖

Cite This Article

KİM, H., PARK, C. M., CHAE, H. D., LEE, S., GOO, J. M. (2015). Impact of radiation dose and iterative reconstruction on pulmonary nodule measurements at chest CT: a phantom study. Diagnostic and Interventional Radiology, Cilt: 21 - Sayı: 6, 459–465. https://ajindex.com/articles/209733

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (13)

  1. Bach PB, Mirkin JN, Oliver TK, et al. Benefits and harms of CT screening for lung cancer: a systemat- ic review. JAMA 2012
  2. [CrossRef] Aberle DR, Adams AM, Berg CD, et al. Reduced lung-cancer mortality with low-dose comput- ed tomographic screening. N Engl J Med 2011
  3. [CrossRef] McCollough CH, Chen GH, Kalender W, et al. Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT. Radiology 2012
  4. [CrossRef] Litmanovich DE, Tack DM, Shahrzad M, Bank- ier AA. Dose reduction in cardiothoracic CT: review of currently available methods. Radio- graphics 2014
  5. [CrossRef] Kubo T, Lin PJ, Stiller W, et al. Radiation dose re- duction in chest CT: a review. AJR Am J Roent- genol 2008
  6. [CrossRef] Willemink MJ, de Jong PA, Leiner T, et al. Iter- ative reconstruction techniques for computed tomography Part 1: technical principles. Eur Radiol 2013
  7. [CrossRef] MacMahon H, Austin JH, Gamsu G, et al. Guide- lines for management of small pulmonary nodules detected on CT scans: a statement from the Fleischner Society. Radiology 2005
  8. [CrossRef] Christe A, Torrente JC, Lin M, et al. CT screening and follow-up of lung nodules: effects of tube current-time setting and nodule size and den- sity on detectability and of tube current-time setting on apparent size. AJR Am J Roentgenol 2011
  9. [CrossRef] Hein PA, Romano VC, Rogalla P, et al. Linear and volume measurements of pulmonary nodules at different CT dose levels - intrascan and inter- scan analysis. Rofo 2009
  10. [CrossRef] Coenen A, Honda O, van der Jagt EJ, Tomiya- ma N. Computer-assisted solid lung nodule 3D volumetry on CT: influence of scan mode and iterative reconstruction: a CT phantom study. Jpn J Radiol 2013
  11. [CrossRef] Willemink MJ, Leiner T, Budde RP, et al. System- atic error in lung nodule volumetry: effect of iterative reconstruction versus filtered back projection at different CT parameters. AJR Am J Roentgenol 2012
  12. [CrossRef] Xie X, Zhao Y, Snijder RA, et al. Sensitivity and accuracy of volumetry of pulmonary nodules on low-dose 16- and 64-row multi-detector CT: an anthropomorphic phantom study. Eur Radi- ol 2013
  13. [CrossRef] Kim H, Park CM, Lee SM, Lee HJ, Goo JM. A com- parison of two commercial volumetry software programs in the analysis of pulmonary ground- glass nodules: segmentation capability and measurem