Journals / International Journal of Medical Biochemistry / 2020 / Cilt: 3 - Sayı: 3

The evaluation of rapid serum tubes for various immunoassay tests

Pages
139–142
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The laboratory turn-around-times for immunoassay tests could be shortened approximately 15-20 minutes by the usage of RSTs.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objectives: One of the most crucial steps on the preanalytical process that affect laboratory test results is proper and timely blood collection.
📊
Key Findings & Data:Methods: Blood samples were collected from 50 healthy volunteers into three blood collection tubes.
💡
Conclusion & Impact:The laboratory turn-around-times for immunoassay tests could be shortened approximately 15-20 minutes by the usage of RSTs.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Objectives: One of the most crucial steps on the preanalytical process that affect laboratory test results is proper and timely blood collection. In this study, we aimed to compare Rapid Serum Tubes (RST) and Serum Separator Tubes (SST) with no additive Z tubes concerning various immunoassay tests. Methods: Blood samples were collected from 50 healthy volunteers into three blood collection tubes. Sera from Z tube, SST and RST were analysed simultaneously for thyroid-stimulating hormone (TSH), free triiodothyronine (fT3), free thyroxine (fT4), folate, vitamin B12, 25-OH vitamin D (25-OHD), parathyroid hormone (PTH), cortisol, ferritin, human chorionic gonadotropin (hCG) on DXI800 autoanalyser (Beckman Coulter, USA). The results were evaluated by comparing each tube pairs (Z tube-SST, Z tube-RST) according to desirable bias. Results: There was no statistically significant difference between the two tube pairs for most analytes. There was not any significant clinical difference according to a desirable bias for the analytes tested. Conclusion: RST offers acceptable clinical performance for the immunoassay tests performed in this study on DXI platform. The laboratory turn-around-times for immunoassay tests could be shortened approximately 15-20 minutes by the usage of RSTs.

🔖

Cite This Article

OĞUZ, E. F., AKBULUT, E. D., ERCAN, M., TOPÇUOĞLU, C., TURHAN, T. (2020). The evaluation of rapid serum tubes for various immunoassay tests. International Journal of Medical Biochemistry, Cilt: 3 - Sayı: 3, 139–142. https://ajindex.com/articles/200282

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (12)

  1. Bowen RA, Remaley AT. Interferences from blood collection tube components on clinical chemistry assays. Biochem Med (Zagreb) 2014;24(1):31–44
  2. Kiechle FL, Betsou F, Blakeney J, Calam RR, Catalasan IM, Raj P, et al. Procedures for the handling and processing of blood specimens for common laboratory tests. Approved Guideline. Clinical Laboratory and Standards Institue 2010;30(10):H18–A4
  3. Bonini P, Plebani M, Ceriotti F, Rubboli F. Errors in laboratory medicine. Clin Chem 2002;48(5):691–8
  4. Carraro P, Plebani M. Errors in a stat laboratory: types and frequencies 10 years later. Clin Chem 2007;53(7):1338–42
  5. Plebani M, Sciacovelli L, Aita A, Padoan A, Chiozza ML. Quality indicators to detect pre-analytical errors in laboratory testing. Clin Chim Acta 2014;432:44–8
  6. Kocijancic M, Cargonja J, Delic-Knezevic A. Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study. Biochem Med (Zagreb) 2014;24(3):368–75
  7. Chance J, Berube J, Vandersmissen M, Blanckaert N. Evaluation of the BD Vacutainer PST II blood collection tube for special chemistry analytes. Clin Chem Lab Med 2009;47(3):358–61
  8. CLSI. Collection of Diagnostic Venous Blood Specimens. 7th ed. CLSI standard GP41. Wayne, PA: Clinical and Laboratory Standards Institute; 2017
  9. Yan R, Lou A, Watts G, Tarr H, Smith H, Kinney L, et al. Comparison of Becton Dickinson Vacutainer rapid serum tube with the serum separator tube for routine chemistry and immunoassay tests. J Clin Pathol 2014;67(7):599–604
  10. Strathmann FG, Ka MM, Rainey PM, Baird GS. Use of the BD vacutainer rapid serum tube reduces false-positive results for selected beckman coulter Unicel DxI immunoassays. Am J Clin Pathol 2011;136(2):325–9
  11. Kristensen GB, Rustad P, Berg JP, Aakre KM. Analytical Bias Exceeding Desirable Quality Goal in 4 out of 5 Common Immunoassays: Results of a Native Single Serum Sample External Quality Assessment Program for Cobalamin, Folate, Ferritin, Thyroid-Stimulating Hormone, and Free T4 Analyses. Clin Chem 2016;62(9):1255–63
  12. Viljoen A, Singh DK, Farrington K, Twomey PJ. Analytical quality goals for 25-vitamin D based on biological variation. J Clin Lab Anal 2011;25(2):130–3.