Koruyucu Halkası Sağlamken Çıkık Gelişen Kısıtlayıcı Total Kalça Protezi Olgusu

Failure of Constrained Total Hip Arthroplasty with Intact Reinforcement Ring

Pages
445–447
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Bu yüzden doğru yerleşim açısının daha geniş bir aralıkta olması uygun olacaktır.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Kalça çıkığı total kalça protezi (TKP) sonrası görülen en sık komplikasyonlardan biridir.
📊
Key Findings & Data:Bu istenmeyen komplikasyonlarla karşılaşmamak için çıkık riskinin artmış olduğu hastalarda sıklıkla kısıtlayıcı sistemler kullanılmaktadır.
💡
Conclusion & Impact:Bu yüzden doğru yerleşim açısının daha geniş bir aralıkta olması uygun olacaktır.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Kalça çıkığı total kalça protezi (TKP) sonrası görülen en sık komplikasyonlardan biridir. Bu istenmeyen komplikasyonlarla karşılaşmamak için çıkık riskinin artmış olduğu hastalarda sıklıkla kısıtlayıcı sistemler kullanılmaktadır. Kısıtlayıcı sistem ile TKP yapılan ve sonrasında koruyucu halkası sağlam olduğu halde kalça çıkığı gelişen iki hasta olgu sunumu olarak paylaşıldı. Hastalardan bir tanesi revizyon ameliyatını kabul etmedi, diğeri ise kliniğimizde opere edildi. Opere edilen hastada kısıtlayıcı liner malrotasyonuna bağlı sıkışma ve liner aşınması gözlendi. Kısıtlayıcı linerin uygun yerleştirilmesi bu ameliyatlarda hayati öneme sahip, aksi takdirde liner aşınması ve kalça çıkığı ile karşılaşılabilir. İmplant üreticisinin yönergelerine uymak gereklidir. Ancak bu kadar küçük bir yerleştirme hatasının bu kadar büyük sonuçları olması implantın bir dezavantajı olarak görülebilir. Zira ameliyat sırasında uygun polietilen liner rotasyonunu belirlemek oldukça zor olabilir. Bu yüzden doğru yerleşim açısının daha geniş bir aralıkta olması uygun olacaktır.

Özet

One of the most common complications after total hip arthroplasty (THA) is dislocation. To avoid this unpleasant complication, we use constrained systems especially in patients at high risk of dislocation. We report two cases of failure of constrained hip system with intact reinforcement ring. We used Trilogy® revision cup in both patients. One of the patients refused surgery. Other patient underwent surgery and we observed abrasion of constrained liner because of impingement. Malrotation of the constrained liner appears to cause impingement. When using constrained liners, caution should be exercised against the risk of impingement. Proper assessment of the constrained liner is mandatory. Otherwise, abrasion of the liner is not a surprise. Instructions of the manufacturer should be followed. The fact that such a small misplacement has such great consequences can also be accepted as a disadvantage of the product. It may be difficult to determine the proper rotation of the polyethylene liner during surgery. Therefore, it would be appropriate to have a wider range of correct polyethylene placement angles.

🔖

Cite This Article

ATLIHAN, D., GÜNAYDIN, F. (2019). Koruyucu Halkası Sağlamken Çıkık Gelişen Kısıtlayıcı Total Kalça Protezi Olgusu. Haseki Tıp Bülteni, Cilt: 57 - Sayı: 4, 445–447. https://ajindex.com/articles/189585

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (1)

  1. Soong M, Rubash HE, Macaulay W. Dislocation after total hip arthroplasty. J Am Acad Orthop Surg 2004;12:314 2. Paterno SA, Lachiewicz PF, Delley SS. The influence of patient-related factors and the position of the acetabular component on the rate of dislocation after total hip replacement. J Bone Joint Surg 1997;79-A:1202. 3. Su EP, Pellicci PM. The role of constrained liners in total hip arthroplasty. Clin Orthop 2004;420:122-9. 4. Lachiewicz PL, Delley SS. The use of constained components in total hip arthroplasty. J Am Acad Orthop Surg 2002;10:4. 5. Kaper BP, Bernini PM, et al. Failure of a constrained acetabular prosthesis of a total hip arthroplasty:a report of four cases. J Bone Joint Surg 1998;80:561-5. 6. Thoms RJ, Marwin SE. A unique failure mechanism of a constrained total hip arthroplasty. J Arthroplasty 2008;23:293-8. 7. Tufescu TV, Dust W. Failure of a new constrained acetabular insert. J arthroplasty 2004;19:238-9.