AI Summary & Key Takeaways
AI InsightThe selectivity coefficients of some anions were calculated by the matched potential method.
📌 3 Key Takeaways
Abstract
A hydrogen ion-selective electrode was prepared using 5,11,17,23-tetra-tert-butyl-25,26,27,28-tetracyanometoxy-calix[4]arene and the possibility of its use as a perchlorate ion-selective electrode was investigated using its characteristic of becoming perchlorate sensitive in acidic regions. The electrode of the optimum characteristic had a composition of 1% ionophore, 66% o-NPOE, and 33% PVC. This electrode exhibited a linear response over the range 1.0 \times 10-1-1.0 \times 10-5 M of perchlorate with a slope of 48.7 \pm 0.5 mV per p[ClO4]. The effects of the pH and the membrane composition were also investigated. The lifetime of the electrode was at least 4 months and its response time was 10-15 s. The selectivity coefficients of some anions were calculated by the matched potential method.
Özet
A hydrogen ion-selective electrode was prepared using 5,11,17,23-tetra-tert-butyl-25,26,27,28-tetracyanometoxy-calix[4]arene and the possibility of its use as a perchlorate ion-selective electrode was investigated using its characteristic of becoming perchlorate sensitive in acidic regions. The electrode of the optimum characteristic had a composition of 1% ionophore, 66% o-NPOE, and 33% PVC. This electrode exhibited a linear response over the range 1.0 \times 10-1-1.0 \times 10-5 M of perchlorate with a slope of 48.7 \pm 0.5 mV per p[ClO4]. The effects of the pH and the membrane composition were also investigated. The lifetime of the electrode was at least 4 months and its response time was 10-15 s. The selectivity coefficients of some anions were calculated by the matched potential method.