Journals / Diagnostic and Interventional Radiology / 2018 / Cilt: 24 - Sayı: 4

Size-specific dose estimates in chest, abdomen, and pelvis CT examinations of pediatric patients

Pages
243–248
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

CONCLUSION In pediatric CT exams involving CAP region, CTDIvol and the water equivalent diameter at the middle of the scan range can be used to provide a reasonable estimate of mean SSDE with an RMSD of 11% at most.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:PURPOSE Size-specific dose estimates (SSDEs) are the latest focus of interest among medical physicists studying radiation dose to the patient in computed tomography (CT).
📊
Key Findings & Data:METHODS In this retrospective study, we examined 105 consecutive pediatric CT exams of chest, abdomen, and pelvis (CAP) classified in 3 separate groups according to age: 0–5 years, 6–10 years, and 11–16 years.
💡
Conclusion & Impact:CONCLUSION In pediatric CT exams involving CAP region, CTDIvol and the water equivalent diameter at the middle of the scan range can be used to provide a reasonable estimate of mean SSDE with an RMSD of 11% at most.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

PURPOSE Size-specific dose estimates (SSDEs) are the latest focus of interest among medical physicists studying radiation dose to the patient in computed tomography (CT). This study aims to make conversions from CTDIvol to SSDE and investigate the relationship between mean SSDE (SSDE) and central SSDE (SSDEcenter) values for pediatric patients of different age groups undergoing chest, abdomen, and pelvis scans. METHODS In this retrospective study, we examined 105 consecutive pediatric CT exams of chest, abdomen, and pelvis (CAP) classified in 3 separate groups according to age: 0–5 years, 6–10 years, and 11–16 years. A MATLAB program was developed to determine SSDE values for each patient along 6 subregions: chest, abdomen, pelvis, chest and abdomen, abdomen and pelvis, and CAP. SSDE values derived over the slice at the center of each scan range (SSDEcenter) were also recorded. SSDE and corresponding SSDEcenter results were compared for each age group. RESULTS Root mean square differences (RMSD) between SSDE and SSDEcenter values ranged between 0.13 mGy and 2.1 mGy through all groups and subregions, corresponding to 1.2% and 11%, respectively. CONCLUSION In pediatric CT exams involving CAP region, CTDIvol and the water equivalent diameter at the middle of the scan range can be used to provide a reasonable estimate of mean SSDE with an RMSD of 11% at most.

🔖

Cite This Article

ÖZSOYKAL, İ., YURT, A., AKGÜNGÖR, K. (2018). Size-specific dose estimates in chest, abdomen, and pelvis CT examinations of pediatric patients. Diagnostic and Interventional Radiology, Cilt: 24 - Sayı: 4, 243–248. https://ajindex.com/articles/173329

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (9)

  1. Shope TB, Gagne RM, Johnson GC. A method for describing the doses delivered by transmission x-ray computed tomography. Med Phys 1981
  2. 488–495. [CrossRef] 2. McCollough CH, Leng S, Yu L, Cody DD, Boone JM, McNitt-Gray MF. CT dose index and patient dose: they are not the same thing. Radiology 2011
  3. 311–316. [CrossRef] 3. International Commission on Radiation Units and Measurements. Report 87: radiation dose and image-quality assessment in computed tomography. 2012. 4. Khatonabadi M, Kim HJ, Lu P, et al. The feasibility of a regional CTDIvol to estimate organ dose from tube current modulated CT exams. Med Phys 2013
  4. 051903. [CrossRef] 5. Christner JA, Braun NN, Jacobsen MC, Carter RE, Kofler JM, Mccollough CH. Size-specific dose estimates for adult patients at CT of the torso 1. Radiology 2012
  5. 841–847. [CrossRef] 6. Menke J. Comparison of different body size parameters for individual dose adaptation in body CT of adults. Radiology 2005
  6. 565–571. [CrossRef] 7. American Association of Physicists in Medicine. Report 204: size specific dose estimates (SSDE) in pediatric and adult body CT examinations. 2011. 8. American Association of Physicists in Medicine. Report 220: use of water equivalent diameter for calculating patient size and size-specific dose estimates (SSDE) in CT. 2014. 9. Wang J, Duan X, Christner JA, Leng S, Yu L, McCollough CH. Attenuation-based estimation of patient size for the purpose of size specific dose estimation in CT. Part I. development and validation of methods using the CT image. Med Phys 2012
  7. 6764–6771. [CrossRef] 10. Wang J, Christner JA, Duan X, Leng S, Yu L, McCollough CH. Attenuation-based estimation of patient size for the purpose of size specific dose estimation in CT. Part II. Implementation on abdomen and thorax phantoms using cross sectional CT images and scanned projection radiograph images. Med Phys 2012
  8. 6772–6778. [CrossRef] 11. Leng S, Shiung M, Duan X, Yu L, Zhang Y. Size-specific dose estimates in chest, abdomen, and pelvis CT: Impact of intra-patient variability in water equivalent diameter. Radiology 2015
  9. 184–190. [CrossRef] 12. International Atomic Energy Association. Human Health Series No 19: Quality assurance programme for computed tomography: Diagnostic and therapy applications. 2012.