Comparison of small-volume tubes and vacuum blood tubes for complete blood count

Pages
24–29
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The design of SVTs may have an important and detrimental effect on the accuracy and efficacy of the results of a CBC.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objectives: A complete blood count (CBC) is one of the most commonly requested clinical laboratory tests.
📊
Key Findings & Data:Vacuum blood tubes are used routinely, and now, new small-volume tubes (SVTs) containing dipotassium ethylenediaminetetraacetic acid (K2EDTA) are also in use.
💡
Conclusion & Impact:The design of SVTs may have an important and detrimental effect on the accuracy and efficacy of the results of a CBC.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Objectives: A complete blood count (CBC) is one of the most commonly requested clinical laboratory tests. Vacuum blood tubes are used routinely, and now, new small-volume tubes (SVTs) containing dipotassium ethylenediaminetetraacetic acid (K2EDTA) are also in use. The aim of this research was to compare SVTs with vacuum blood tubes for use in a CBC. Methods: Venous blood samples were taken from 40 healthy volunteers and were collected in BD Vacutainer (Becton, Dickinson and Company, Franklin Lakes, NJ, USA) K2EDTA tubes and BD Microtainer (Becton, Dickinson and Company, Franklin Lakes, NJ, USA) K2EDTA tubes. CBC parameters were analyzed using an ABX Pentra DF 120 device (Horiba, Ltd., Kyoto, Japan). Results: Red blood cells (RBC), hemoglobin (HGB), hematocrit (HCT), mean corpuscular hemoglobin (MCH), mean corpuscular hemoglobin concentration (MCHC), and basophil (BASO) levels were found to be statistically significantly higher, while platelet (PLT) levels were determined to be statistically significantly lower in the SVT analyses compared with those of the vacuum blood tubes. When the percentage difference was compared with the total allowable error, RBC, HGB, HCT, MCH, MCHC, red cell distribution width, white blood cell count, neutrophil, lymphocyte, monocyte, eosinophil, and BASO values demonstrated a general trend of positive bias, while PLT values demonstrated a general trend of negative bias on a Bland-Altman bias plot. Conclusion: The results of this study demonstrated that venous blood samples from SVTs generated a significant bias in the CBC results. In addition, clotted samples were observed more frequently among those collected in SVTs in this study. The design of SVTs may have an important and detrimental effect on the accuracy and efficacy of the results of a CBC.

🔖

Cite This Article

ÜNAL, K. (2019). Comparison of small-volume tubes and vacuum blood tubes for complete blood count. International Journal of Medical Biochemistry, Cilt: 2 - Sayı: 1, 24–29. https://ajindex.com/articles/167212

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (15)

  1. Buttarello M, Plebani M. Automated blood cell counts: state of the art. Am J Clin Pathol 2008;130:104–16
  2. Narayanan S. The preanalytic phase. An important component of laboratory medicine. Am J Clin Pathol 2000;113:429–52
  3. Clinical and Laboratory Standars Institute. GP39-A6 Tubes additives for venous and capillary blood specimen collection; Approved Standard. 6th ed. Wayne: Clinical and Laboratory Standards Institute; 2010
  4. Banfi G, Salvagno GL, Lippi G. The role of ethylenediamine tetraacetic acid (EDTA) as in vitro anticoagulant for diagnostic purposes. Clin Chem Lab Med 2007;45:565–76
  5. Bowen RA, Remaley AT. Interferences from blood collection tube components on clinical chemistry assays. Biochem Med (Zagreb) 2014;24:31–44
  6. Clinical and Laboratory Standars Institute. GP44-A4 Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests; Approved Guideline. Wayne: Clinical and Labaratory Standards Institute; 2010
  7. Bowen RA, Hortin GL, Csako G, Otañez OH, Remaley AT. Impact of blood collection devices on clinical chemistry assays. Clin Biochem 2010;43:4–25
  8. Sanchez-Giron F, Alvarez-Mora F. Reduction of blood loss from laboratory testing in hospitalized adult patients using small-volume (pediatric) tubes. Arch Pathol Lab Med 2008;132:1916–9
  9. Ricós C, Alvarez V, Cava F, García-Lario JV, Hernández A, Jiménez CV, et al. Current databases on biological variation: pros, cons and progress. Scand J Clin Lab Invest 1999;59:491–500
  10. Evans JD. Straightforward statistics for the behavioral sciences. Pacific Grove: Brooks/Cole; 1996
  11. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–10
  12. Krouwer JS, Tholen DW, Garber CC, Goldschmidt HM, Kroll MH, et al. Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline—2nd ed. NCCLS document EP9-A2. Wayne: Clinical and Laboratory Standards Institute; 2002
  13. Inal S, Kelleci M. Relief of pain during blood specimen collection in pediatric patients. MCN Am J Matern Child Nurs 2012;37:339–45
  14. Smoller BR, Kruskall MS, Horowitz GL. Reducing adult phlebotomy blood loss with the use of pediatric-sized blood collection tubes. Am J Clin Pathol 1989;91:701–3
  15. Xu M, Robbe VA, Jack RM, Rutledge JC. Under-filled blood collection tubes containing K2EDTA as anticoagulant are acceptable for automated complete blood counts, white blood cell differential, and reticulocyte c