Electric Pulp Testing with Examination Gloves

Umut Zengin(İSTANBUL ÜNİVERSİTESİ, DİŞ HEKİMLİĞİ FAKÜLTESİ, DİŞ HEKİMLİĞİ (DR))0009-0004-1703-1900
Melek Tufan Yücesoy(İSTANBUL ÜNİVERSİTESİ, DİŞ HEKİMLİĞİ FAKÜLTESİ, DİŞ HEKİMLİĞİ (DR))0009-0002-3603-9989
Pages
—
Publication Date
2023-12-20
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Conclusion In conclusion, the clinician should not use the EPT while wearing examination gloves.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Aim The objective of this study is to determine differences between EPT results for the same teeth with the operator’s hands being gloved and ungloved.
📊
Key Findings & Data:Results Statistically significant differences between EPT results for the same teeth with the operator’s hands being gloved and ungloved were observed.
💡
Conclusion & Impact:Conclusion In conclusion, the clinician should not use the EPT while wearing examination gloves.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Dental pulp tests · Electric pulp testing · Examination gloves

Abstract

Aim The objective of this study is to determine differences between EPT results for the same teeth with the operator’s hands being gloved and ungloved. Material and method Electric pulp tests were conducted for the same teeth with the operator’s hands being gloved and ungloved. Twenty male and twenty female patients were included in this study. Results Statistically significant differences between EPT results for the same teeth with the operator’s hands being gloved and ungloved were observed. Conclusion In conclusion, the clinician should not use the EPT while wearing examination gloves.

🔖

Cite This Article

Zengin, U., Yücesoy, M. T., Moghbel, M. (2023). Electric Pulp Testing with Examination Gloves. Eurasian Dental Research, Volume: 1 Issue: 3. https://ajindex.com/articles/1645123

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (5)

  1. Chen E, Abbott P V. Dental Pulp Testing: A Review. Int J Dent. 2009;2009:1-12. doi:10.1155/2009/365785
  2. Jafarzadeh H, Abbott P V. Review of pulp sensibility tests. Part II: Electric pulp tests and test cavities. Int Endod J. 2010;43(11):945-958. doi:10.1111/j.1365-2591.2010.01760.x
  3. Lin J, Chandler NP. Electric pulp testing: A review. Int Endod J. 2008;41(5):365-374. doi:10.1111/j.1365-2591.2008.01375.x
  4. Samuel S, Thomas A, Singh N. A comparative study of pulse oximetry with the conventional pulp testing methods to assess vitality in immature and mature permanent maxillary incisors. CHRISMED J Heal Res. 2014;1(4):235. doi:10.4103/2348-3334.142985
  5. Sharma A, Madan M, Shahi P, Sood P, Shahi N. Comparative Study of Pulp Vitality in Primary and Young Permanent Molars in Human Children with Pulse Oximeter and Electric Pulp Tester. Int J Clin Pediatr Dent. 2015;8(2):94-98. doi:10.5005/jp-journals-10005-1291