Unexpected laboratory results in cold agglutinin disease

Pages
40–43
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The automated hematology analyzer results indicated a very low hematocrit (7%) and erythrocyte count (0.57x106 mm³).

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Cold agglutinin disease (CAD) is an autoimmune hemolytic disease in which antibodies for erythrocyte surface antigens are activated by low temperatures, causing agglutination of erythrocytes.
📊
Key Findings & Data:In this case, unexpected haptoglobin (Hp) and glycated hemoglobin (HbA1c) results, as well as complete blood count (CBC), were reported.
💡
Conclusion & Impact:The automated hematology analyzer results indicated a very low hematocrit (7%) and erythrocyte count (0.57x106 mm³).
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Cold agglutinin disease (CAD) is an autoimmune hemolytic disease in which antibodies for erythrocyte surface antigens are activated by low temperatures, causing agglutination of erythrocytes. Most of the published cases in the literature are related to the clinical presentation of the disease. Cold agglutinins can also interfere with laboratory tests. In this case, unexpected haptoglobin (Hp) and glycated hemoglobin (HbA1c) results, as well as complete blood count (CBC), were reported. A 55-year-old woman with joint paint presented at the rheumatology polyclinic. Blood samples were collected into anticoagulated and gel separation tubes for CBC, HbA1c, and biochemistry tests. The automated hematology analyzer results indicated a very low hematocrit (7%) and erythrocyte count (0.57x106 mm³). The erythrocyte indices, platelet count, and mean platelet volume could not be measured. Hp was undetectable. The HbA1c value was below the reference threshold. The patient’s file revealed that she had been diagnosed with CAD 5 years earlier. The samples were reanalyzed after warming in a Benmarie for 25 minutes to 37°C. While the CBC results improved, the HbA1c and Hp results did not change. Informing the laboratory when samples to be used are from patients with CAD and providing the proper temperature conditions during transportation will ensure accurate results, as well as decrease workload and laboratory costs.

🔖

Cite This Article

YASAR, N. E., OZGENC, A., BOLAYIRLI, I. M., ADIGUZEL, M., KONUKOĞLU, D. (2018). Unexpected laboratory results in cold agglutinin disease. International Journal of Medical Biochemistry, Cilt: 1 - Sayı: 1, 40–43. https://ajindex.com/articles/161628

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (11)

  1. Swiecicki PL, Hegerova LT, Gertz MA. Cold agglutinin disease. Blood 2013;122:1114–21
  2. Ercan S, Calışkan M, Koptur E. 70-year old female patient with mismatch between hematocrit and hemoglobin values: the effects of cold agglutinin on complete blood count. Biochem Med (Zagreb) 2014;24:391–5
  3. Javed R, Datta SS, Basu S, Chakrapani A. Resolution of Serologic Problems Due to Cold Agglutinins in Chronic Lymphocytic Leukemia. Indian J Hematol Blood Transfus 2016;32:290– 3
  4. Kalyani R, Thej MJ, Thomas AK, Raveesha A. Chronic Cold Agglutinin Disease: A Case Report with Review of Literature. Journal of Clinical and Diagnostic Research 2012;6:480–2
  5. Nikousefat Z, Javdani M, Hashemnia M, Haratyan A, Jalili A. Cold Agglutinin Disease; A Laboratory Challenge. Iran Red Crescent Med J 2015;17:e18954
  6. Kumar TB, Bhardwaj N. Platelet cold agglutinins and thrombocytopenia: A diagnostic dilemma in the intensive care unit. J Anaesthesiol Clin Pharmacol 2014;30:89–90
  7. Lodi G, Resca D, Reverberi R. Fatal cold agglutinin-induced haemolytic anaemia: a case report. J Med Case Rep 2010;4:252
  8. De Albuquerque Wobeto VP, Zaccariotto TR, de Fátima Sonati M. Polymorphism of human haptoglobin and its clinical importance. Genet Mol Biol 2008;31:602–20
  9. Van Rijn HJ, van der Wilt W, Stroes JW, Schrijver J. Is the turbidimetric immunoassay of haptoglobin phenotype-dependent? Clin Biochem 1987;20:245–8
  10. Cavagnolli G, Pimentel AL, Freitas PA, Gross JL, Camargo JL. Factors affecting A1C in non-diabetic individuals: Review and meta-analysis. Clin Chim Acta 2015;445:107–14
  11. Rhea JM, Molinaro R. Pathology consultation on HbA(1c) methods and interferences. Am J Clin Pathol 2014;141:5–16.