Comparison of stability of seven biochemistry analytes in serum separator and plasma tubes

Pages
14–18
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The routine chemistry analytes studied were within the clinically acceptable range in both plasma and serum specimens, but validation and estimation of reference intervals for plasma tubes are needed.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objectives: The aim of the present study was to compare the stability of 7 routine analytes using a BD Vacutainer Barricor plasma blood collection tube, a BD Vacutainer SST (Becton Dickinson and Company, Franklin Lakes, NJ, USA) and a Vacusera SST ((Disera A.S., Izmir, Turkey).
📊
Key Findings & Data:Methods: Five mL blood samples from 60 volunteers were collected into the 3 tubes.
💡
Conclusion & Impact:The routine chemistry analytes studied were within the clinically acceptable range in both plasma and serum specimens, but validation and estimation of reference intervals for plasma tubes are needed.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Objectives: The aim of the present study was to compare the stability of 7 routine analytes using a BD Vacutainer Barricor plasma blood collection tube, a BD Vacutainer SST (Becton Dickinson and Company, Franklin Lakes, NJ, USA) and a Vacusera SST ((Disera A.S., Izmir, Turkey). Methods: Five mL blood samples from 60 volunteers were collected into the 3 tubes. Glucose, urea, creatinine, lactate dehydrogenase (LD), sodium (Na), chloride (Cl) and potassium (K) levels were analyzed at 0, 4, 8, and 12 hours on a Cobas 8000 analyzer ((F. Hoffmann-La Roche Ltd., Basel, Switzerland). Bias and bias% results were calculated and compared to clinically acceptable limits (CAL). Results: Intraclass correlation results revealed excellent agreement in the 7 chemistry analytes between the 3 tubes (0.863-1.000). The CAL for Cl, creatinine, glucose, K, LD, Na and urea are 3 mmol/L, 10%, 0.3 mg/dL, 10%, 0.5 mmol/L, 20%, 3 mmol/L and 2mg/dl, respectively. At the 12th hour, there was a difference observed in glucose between Barricor- SST and Barricor-Vacusera SST were 10.06 mg/dL (11.38 %) and 11.14 mg/dL (-12.76%). Analytical coefficient (CV) of variation results for glucose, LD, creatinine, urea, Na, K, and Cl were 3.22%, 5.80%, 5.28%, 3.90%, 2.00%, 1.95%, and 1.37% respectively. Conclusion: Although the stability of all 7 analytes was within the CAL during a 12-hour period, clinically different results were obtained. In terms of stability evaluation, the use of plasma might be preferable in clinical laboratories. The routine chemistry analytes studied were within the clinically acceptable range in both plasma and serum specimens, but validation and estimation of reference intervals for plasma tubes are needed.

🔖

Cite This Article

AVCI, E., DEMİR, S., NAR, R., AKPİNAR, K. (2020). Comparison of stability of seven biochemistry analytes in serum separator and plasma tubes. International Journal of Medical Biochemistry, Cilt: 3 - Sayı: 1, 14–18. https://ajindex.com/articles/155302

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (9)

  1. Çuhadar S. Preanalytical variables and factors that interfere with the biochemical parameters: a review. OA Biotechnology 2013;2:19
  2. Sciacovelli L, O'Kane M, Skaik YA, Caciagli P, Pellegrini C, Da Rin G, et al, Quality Indicators in Laboratory Medicine: from theory to practice Preliminary data from the IFCC Working Group Project ‘‘Laboratory Errors and Patient Safety’’. Clin Chem Lab Med 2011;49:835–44
  3. Green SF. The cost of poor blood specimen quality and errors in preanalytical processes. Clin Biochem 2013;46:1175–9
  4. Plebani M, Sciacovelli L, Aita A, Chiozza ML. Harmonization of pre-analytical quality indicators. Biochem Med (Zagreb) 2014;24:105–13
  5. Sandberg S, Christensen NG, Horgen E, Jacobsen H, Johansen K. Evaluation of Serum Separator Tubes as a Mail Transport Device in Primary Health Care. Scand J Prim Health Care 1988;6:251–4
  6. Cuhadar S, Atay A, Koseoglu M, Dirican A, Hur A. Stability studies of common biochemical analytes in serum separator tubes with or without gel barrier subjected to various storage conditions. Biochem Med (Zagreb) 2012;22:202–14
  7. Bowen RA, Remaley AT. Interferences from blood collection tube components on clinical chemistry Assays. Biochem Med (Zagreb) 2014;24:31–44
  8. Goce D, Johnston J. Is the BD Barricor Tube the New Standard for Lithium Heparin Plasma?. Developments Clin Med Pathol 2018;1:1–5
  9. Arslan FD, Karakoyun I, Basok BI, Aksit MZ, Baysoy A, Ozturk YK, et al. The local clinical validation of a new lithium heparin tube with a barrier: BD Vacutainer® Barricor LH Plasma tube. Biochem Med (Zagreb) 2017;27:030706.