Objective & Hypothesis:This study examines the dose difference for a variety of phantom materials that can be employed for “Leksell Gamma” Knifes.
📊
Key Findings & Data:Geant4 was utilized in the analysis of the dose distributions for collimator helmet sizes of 4mm and 8mm.
💡
Conclusion & Impact:In addition, the polystyrene and PMMA (phantom is also quite appropriate in terms of evaluating the dose profiles of the Gamma Knife unit.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Geant4 · simulation · gamma knife
Abstract
This study examines the dose difference for a variety of phantom materials that can be employed for “Leksell Gamma” Knifes. These materials resemble human tissue by not including the skull bone. Geant4 was utilized in the analysis of the dose distributions for collimator helmet sizes of 4mm and 8mm. The phantom is illustrated with a radius of 80mm. Water, brain, Poly-methyl methacrylate (PMMA) and polystyrene were being considered as the main material types. Results proved that there is no considerable differences for radiation dosimetries depending on the material types. In addition, the polystyrene and PMMA (phantom is also quite appropriate in terms of evaluating the dose profiles of the Gamma Knife unit.
Dağlı, Ö., Bostancı, G. E., Emmez, H., Kurt, G., Ekinci, F., Güzel, M. S. (2021). Geant4 based dosimetry evaluation for gamma knife using different phantom materials. Communications Faculty of Sciences University of Ankara Series A2-A3 Physical Sciences and Engineering, Volume: 63 Issue: 2. https://doi.org/10.33769/aupse.956400
Bertolino, N., Experimental validation of Monte Carlo simulation of Leksell Gamma Knife Perfexion stereotactic radiosurgery system. PhD Thesis, Università Degli Studi Di Milano, (2009), 6-17.
Goetsch, S.J., Risk analysis of Leksell Gamma Knife Model C with automatic positioning system, Int. J. Radiat. Oncol. Biol. Phys., 52 (2002), 869-877. https://doi.org/:10.1016/s0360-3016(01)02718-3
Leksell Gamma Plan Instructions for Use for Version 4.0-Target Series, Elekta, 1996.
Ekinci, F., Bostanci, E., Dagli, O., Guzel, M.S., Analysis of Bragg curve parameters and lateral straggle for proton and carbon beams, Commun. Fac. Sci. Univ. Ank. Ser. A2-A3, 63 (1) (2021), 32-41. https://doi.org/:10.33769/aupse.864475
Dagli, O., Bostanci, E., Emmez, O.H., Ekinci, F., Celtikci, E., A statistical analysis of the effect of different embolization materials on Gamma Knife arteriovenous malformation dose distributions, arXiv:2101.08034 (2021). https://arxiv.org/abs/2101.08034
Ekinci, F., Bostanci, E., Dagli, O., Guzel, M.S., Recoil analysis for heavy ion beams, arXiv:2101.09720 (2021). https://arxiv.org/abs/2101.09720
ICRU Report 62 Prescribing, Recording and Reporting Photo Beam Therapy (Supplement to ICRU Report 50), Bethesda, (1999).
Aksoy, A., Karsli, O., Aydin, A., Kaya, C., Ketenoglu, B., Ketenoglu, D. and Yavas, O. Current status of Turkish accelerator and radiation laboratory in Ankara: the TARLA facility, Can. J. Phys., 96 (7) (2018), 837-842. https://doi.org/10.1139/cjp-2017-0750
Ketenoglu, D., Bostanci, E., Aydin, A., Ketenoglu, B., A hard X-ray self-amplified spontaneous emission free-electron laser optimization using evolutionary algorithms for dedicated user applications, Turk. J. Phys., 43 (2019), 551-555. https://doi.org/10.3906/fiz-1909-5
Agostinelli, S., Allision, J., Amaka, K., Apostolakis, J., Araujo, H., Arce, P., et al., GEANT4-A simulation toolkit, Nucl. Instrum. Methods Phys. Res. A, 506 (2003), 250- 303. https://doi.org/10.1016/S0168-9002(03)01368-8
Li, J.S., Ma, C.M., A method to reduce the statistical uncertainty caused by high energy cutoffs in Monte Carlo treatment planning, J. Phys. Conf. Ser., 102 (2008), 012015. https://doi.org/:10.1088/1742-6596/102/1/012015
Cuttone, G., Mongelli, V., Romano, F., Russo, G., Sabini, M., Geant4 based simulation of the Leksell Gamma Knife for treatment planning validations, 14th Geant4 Users and Collaboration Workshop, Catania, Italy, (2009).
Dağlı, Ö., Tanır, A.G., Kurt, G., Analysis of radiation dose distribution Inhomogenity effects in Gamma Knife radiosurgery using Geant4, Politeknik Dergisi, (2021). https://doi.org/10.2339/politeknik.674718
Sandilos, P., Tatsis, E., Vlachos, L., Dardoufas, C., Karaiskos, P., Georgiou, E., and Angelopoulos, A., Mechanical and dose delivery accuracy evaluation in radiosurgery using polymer gels, J. Appl. Clin. Med. Phys., 7 (4) (2006), 13-21. https://doi.org/:10.1120/jacmp.v7i4.2273
Cheung, J.Y.C., Yu, K.N., Yu, C.P., Ho, R.T.K., Choice of phantom materials for dosimetry of Leksell Gamma Knife unit: A Monte Carlo study, Med. Phys., 29 (10) (2002), 2260. https://doi.org/:10.1118/1.1508797
Senirkentli, B.G., Ekinci, F., Bostanci, E., Guzel, M.S., Dagli, O., Karim, A., Mishra, A., Proton therapy for Mandibula plate phantom, Healthcare, 9 (2) (2021), 167. https://doi.org/:10.3390/healthcare9020167