Wrongs known as right in thyroid scintigraphy and uptake study

Wrongs known as right in thyroid scintigraphy and uptake study

Sayfa
142–147
DOI
—

Abstract

Objectives: Thyroid scintigraphy using 99mTc-pertechnetate is commonly used to study function and structureof thyroid gland. Pin-hole collimator is generally preferred in thyroid scintigraphy and uptake studies. Thepurpose of the present study was to determine actual radiopharmaceutical uptake value in an experimental99mTc-pertechnetate thyroid scintigraphy and uptake model.Methods: Thyroid hyperactive and hypoactive nodule models were created using 4 mCi (148 MBq) 99mTcpertechnetate. In the experimental model, 4 mm, 6 mm and 8 mm diameter pin-hole collimators, and 5 cm, 7cm and 10 cm object-to-pinhole distances were investigated.Results: In thyroid hyperactive nodule model, despite the same activity value, uptake at 7 cm object-to-pinholedistance was higher compared to 10 cm distance (122% and 103%, respectively). In the patient with Graves’disease, despite the same activity value, uptake at 5 cm object-to-pinhole distance was higher compared to 10cm distance (8% and 4%, respectively). In thyroid hypoactive nodule model, 4 mm, 6 mm and 8 mm diameterspin-hole collimators were imaged at 5 cm, 10 cm and 15 cm object-to-pinhole distances. The resolutiondifferences between the images were evaluated.Conclusion: It was determined that imaging using 10 cm object-to-pinhole distanceand 4 mm diameter pinhole collimator was best in terms of image resolution and optimum 99mTc-pertechnetate uptake level.