Dergiler / Balıkesir Üniversitesi Fen Bilimleri Enstitüsü Dergisi / 2002 / Cilt: 4 - Sayı: 2

Lead(II) and cobalt(III) hepthyldithiocarbamates as new coflotation reagents for preconcentration of cadmium before its ETAAS determination

Lead(II) and cobalt(III) hepthyldithiocarbamates as new coflotation reagents for preconcentration of cadmium before its ETAAS determination

Sayfa
85–89
DOI
—

Abstract

Two collectors, lead(II) hepthyldithiocarbamate, Pb(HPDTC)2, and cobalt(III) hepthyldithiocarbamate, Co(HPDTC)3, for colloidal precipitate flotation of Cd in traces from water matrices were proposed. After flotation Cd was determined by electrothermal atomic absorption spectrometry (ETAAS). The optimal conditions for effective Cd flotation performed separately by each collector were established. The results of ETAAS analyses are compared with those obtained by the method of inductively coupled plasrna-atomic emission spectrometry. The ETAAS limit detection of Cd by Pb(HPDTC)2 is 0.0048 $mu g/l$, while by Co(HPDTC)3 is 0.003 $mu g/l$.